Source |
- 2.5GeV PF Bending magnet (B10)
|
Optics |
- Monochtomator (Fix-exit double crystal Si(111), Indirect water cooling)
- 2 dimensional focusing mirror (Bent cylinder, Rhodium coated Si substrate, 3.5 mrad glancing angle)
- Focal point : On the SAXS detector surface
|
Energy range |
- 7.0 - 14.0 keV (0.89 - 1.77 Å)
|
Resolution (ΔE/E) |
|
Beam size |
- 0.18 (V) mm x 0.63 (H) mm @focal point
|
Photon flux |
- 1.1 x 1011 photons/s (Size-definition slit full open, 1.5 Å)
- 3.8 x 1010 photons/s (Size-definition slit:1.4 x 1.6 mm2, 1.5 Å)
|
Instruments |
- Camera length:0.25, 0.5, 1.0, 2.0, 2.5 m(SAXS)
- Camera length:0.83, 1.1, 1.6, 2.6 m (When using WAXS chamber)
- PD embedded beam stopper/GISAXS beam stopper
- Delay pulse generator
- Chiller for sample temperature adjustment (5 - 70 °C)
- Incubator (3 - 45 °C)
- Block incubator (4 – 100 &Cring;、1.5ml, 0.5 ml tube)
- Microcentrifuge (1.5, 0.5 ml, -20 – 40 °C)
- Perista pump
- Sample solution flow system for mitigating the effect of radiation damage.
- Rotary Sample Changer (36 samples)
- HPLC system for SEC-SAXS
- Microspectrophotometer (~ 2 μL)
【Common instruments:BL-6A、BL-10C、BL-15A2】
- General GISAXS stage
- Heating/Cooling stage:10002L(Linkam scientific、-190 - 600 °C), HSC302-LN190(Instec、-190 - 400 °C)
- One-dimensional tensile tester (ISUT-2207、IS GIKEN)
- DSC stage(AA1712-1303、Rigaku)
|
Detector |
- SAXS : PILATUS3 2M(DECTRIS AG)
- WAXS : PILATUS3 200K(DECTRIS AG)
|
Remote / Mail-in |
- Remote experiment using some instruments is available (Please consult with beamline contact person).
|
Special notes |
- A flat card detector (C9728DK-10, Hamamatsu Photonics) is also available as a WAXS detector.
|